Profibus Interface PR 5210 Operating Manual
66 Sartorius
6.3.3.20 Parameter P52: Weight & Measuresmode
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 W&M
W&M mode 0 = off, 1= on = OIML, has to be set before start of calibration
6.3.3.21 Parameter P53: A/D converter sample time (measuring rate)
Read only
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 SAMPLETIME
SAMPLETIME 0: 6.25 Hz == 160 ms = default
1: 12.5 Hz == 80 ms
2: 25 Hz == 40 ms
3: 50 Hz == 20 ms
4: 100 Hz == 10 ms
6.3.4 Parameter P99: Access code
Write only
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 CODE MSB
Byte 1 ""
Byte 2 ""
Byte 3 CODE LSB
CODE If an access code was entered in the configuration tool, parameters and the registers for limits
can only be written if the access code is contained in register P99. Parameter P99 can always be
written. To activate the protection P99 has to be set to –1 again. If CODE is set to 0, no access check is
done.