PR 5210 Operating Manual Technical Data
Sartorius 69
7.9 Environmental effects
7.9.1 Environmental conditions
Temperature range
Reference temperature 23 °C
Ambient temperature operation -10... +55 °C ( -10°C... +40 °C for 'legal for trade' application)
Switch-on temperature 0... +55 °C ( 0 °C... +40 °C for 'legal for trade' application)
Storage/ transport -40... +70 °C
Humidity < 95 %, without condensation, (acc. to IEC 68-2)
Protection type IP 20
Vibration to IEC 68-2-6, test Fc
7.9.2 Electromagnetic compatibility (EMC)
All data comply to NAMUR NE 21 and EN 45501
Housing Radio frequency electromagnetic field
(80 – 1000 MHz)
EN 61000-4-3 10 V/m
Radio frequency electromagnetic field
(900 MHz pulse mod.)
EN 61000-4-3 10 V/m
Electrostatic discharge (ESD) EN 61000-4-2 6/8 kV
Signal and control lines Electrical fast transients (Burst) EN 61000-4-4 1 kV
Peak voltage (surge) 1.2/50 µs EN 61000-4-5 0.5/1 kV
Conducted disturbances by radio-
frequency (0.15 – 80 MHz)
EN 61000-4-6 10 V
Conducted common mode disturbance
(0 – 150 kHz)
EN 61000-4-16 10 V
DC input power port Electrical fast transients (Burst) EN 61000-4-4 2 kV
Peak voltage (surge) 1.2/50 µs EN 61000-4-5 0.5/1 kV
Conducted disturbances by radio-
frequency (0.15 – 80 MHz)
EN 61000-4-6 10 V
Conducted common mode disturbance
(0 – 150 kHz)
EN 61000-4-16 10 V
Ripple on DC input voltage EN 61000-4-17 +20 / -15%.
5% ripple
Voltage variations EN 61000-4-11 40% / 0%
Voltage dips EN 61000-4-11 20 ms
7.9.3 RF interference suppression
Electromagnetic emission acc. To EN 55011 group 1, limit value class B