13
SARTORIUS QC - Scales
Adjusting the Upper Middle Stop
Remove the weighing pan.
There must not be any load on the scale when the upper middle stop is
being adjusted.
Adjust the upper middle stop so that the gap between the weighing pan
and the stop is 0.6 mm (tolerance range: + 0.1 mm); check the gap size
with a feeler gauge.
When the gap is correct, place the weighing pan on the scale.
Adjusting the Corner Stops
Remove the weighing pan and place it upside down on the scale.
Place a load equal to the maximum capacity in Position 1 on the
weighing pan.
Use a feeler gauge to check the distance between the corner stop screws
(115) and the weighing pan.
Proceed as in the following example to adjust the corner stops:
The actual value of the gap (as determined with the feeler gauge) is
1mm.
The target value is 0.25 mm.
The deviation (actual minus target value) is, in this example, 0.75mm.
The lead of the stop screws is 1 mm, i.e.:
1 turn clockwise increases the gap by 1 mm
1 turn counter-clockwise decreases the gap by 1 mm
The adjustment step is calculated from the above values:
1 turn of the stop screw changes the gap size by 1 mm, i.e. in the above
example, you would need to turn the screw ¾ of a turn counter clockwise
to achieve the required adjustment (deviation: 0.75 mm).
Before adjusting the screws (115), remove the weighing pan from the
scale.
Proceed in the same manner to adjust the corner stops in positions 2 – 4.
After completing this adjustment procedure, place the weighing pan right
side up on the scale.
When you have finished adjusting the corner stops, proceed to the
adjustment/calibration functions.