Counting
Counting u/w key Code  2  1 4*
Symbol displayed:  P
The counting program allows automatic conversion of weights into piece counts
based on a reference sample quantity and weight.  A weight readout is stored as 
a reference sample quantity (factory setting: 10 pcs = pieces).
General Functions in the Counting Program
Function/Application Entry Mode User Data
Guide Output
Display ID
Clear the counting  c/c key + CF –
application function u/w key
Display information: i/i key + reF –
reference sample  u/w key
quantity “nRef”or average 
piece weight “wRef”
Data output:   i/i key + nRef
reference sample quantity  p/p key + reF wRef
“nRef” or average  u/w key
piece weight “wRef”
With menu setting  7 1 2:  
automatic!
Entry of an average  Numeric keys 0 –9
piece weight “wRef” and . + u/w key – –
Entry of a reference sample  Numeric keys 0–9
quantity “nRef” without . + ––
u/w key
A number entered manually without a decimal point is always a piece count!
A number entered manually with a decimal point is always a weight value!
* = factory setting
3–4