24
Test Parameters
Enter external connections
Standard test: Yes|No
In the menu “Test parameters: Basic settings,” parameters of the configuration are
entered. The function keys are used there to enter the appropriate settings.
The parameters for the standard test are defined in the configured Service Menu.
Unlike the defaults (standard values), the values can be individually adjusted.
For the diffusion test, you have the following options:
Pressure unit: mbar|psi|hPa
Automatic test mode: No|Yes
Enter parameters for the diffusion test
Test pressure: from 50-8000 mbar
Stabilization time: in min.
Test time: in min.
Max. diff.: in ml/min
Net volume: in ml
In the menu “Test parameters: diffusion test,“ the parameters for the diffusion test
are entered with or without net volume measurement.
For tests with volume measurements:
– Test pressure
– Stabilization time
– Test time
– Max. diffusion
When conducting tests without volume measurement, net volume should be
entered separately.
The test parameter data can be found in the validation or data sheets documents of
the respective filter cartridge.
Locks
The locks are activated for the test parameters (default). After entering with
the Shift + F keys, the programmer can lock the data to prevent any changes.
The dialog boxes are then highlighted in gray. The user can no longer change
the entered data.
Enter record data 1
Company: Company name
Building: Building name; house number
Department: Identification of the department
Production site: Production sector
Product: Product name
Product lot: Batch number
Filter: Filter size
Filter lot: Number of the filter batch
Fig. 7-8 | Test parameters:
External connections
Fig. 7-9 | Test parameters:
Diffusion test
Fig. 7-10 | Test program: Log data 1