Operation 49
When loading the test results from SD card or the network, all test results located
on the SD card or network are displayed in a list. This list has the same structure
as the list of the internal test results. Here you can highlight individual results or
result blocks and display or print them out.
It is not possible to save external result files back into the Sartocheck
®
4 plus
Filter tester results database. It is only possible to display the external result files.
7.18 Special Functions In the menu “Special functions”, you will find Sartocheck
®
4 plus Filter tester
functions that are (F6 – Main Menu) not directly linked to bag or filter testing.
Here, the function tests for the management unit and the test unit are carried out.
The cleaning procedure and drying procedure for the internal pneumatic block
and internal volume reference tank can be selected here. Furthermore, pressure
gauge functions can be activated in the special functions menu. The user can change
their own password as long as they have access rights to this menu.
For more details, refer to the menu “Service user management”.
In the menu “Special functions”, the corresponding functions can be activated
with the F1-F6 keys.
While the Special Functions menu is active, visualizing a currently running test is not
possible. Any test unit that is currently active will not be selectable through the
Special Functions menu.
7.18.1 Function Test MU (F1) With the Function test MU (management unit), you can check Sartocheck
®
4 plus
Filter tester peripherals.
F1: Printer function test
Prints out all print characters in black and red on the internal printer
F2: SD card function test (SD card reader)
After inserting an SD card, you can select from three different test procedures:
Read only|Check file format|Write and read
F3: Touch screen function test
The function of each individual touch field can be checked
F5: Format SD card
Fig. 7-51 | Special functions
Fig. 7-52 | Function test MU