64 Operation
During calibration, the results are displayed as a deviation from the
adjustment curves.
Sartocheck
®
4 plus Bag tester:
If the deviation is below 0.025 % FS, an adjustment is not required.
At deviations of 0.025 – 0.5 % FS, an adjustment is recommended.
At deviations > 0.5 % FS, an adjustment is mandatory.
Sartocheck
®
4 plus Filter tester:
If the deviation is below 0.05 % FS, an adjustment is not required.
At deviations of 0.05 – 0.1 % FS, an adjustment is recommended.
At deviations > 0.1 % FS, an adjustment is mandatory.
During adjustment, a new adjustment curve is calculated and activated using
the entered values. The data of the adjustment can be printed out on the internal
printer after the adjustment is completed. The new adjustment curve can then
be checked with a calibration.
“Page up” takes you back to the selection screen for test units (MultiUnits) and
sensors. Depending on “F10 perform calibration | adjustment?”, “Page down” is used
to immediately start an additional calibration or adjustment.
Fig. 7-87 | Result of the calibration
Fig. 7-88 | Result of the adjustment