Troubleshooting 71
Error-List
Legend for ‘Type’
‘R’ Commit severity ‘Remark’
‘W’ Commit severity ‘Warning’
‘E’ Commit severity ‘Error’
Placeholder in ‘Usertext’
<ARG1>, <ARG2>, <ARG3> are placeholder that are filled with variable information like filename, testprogramnumber, testunit-
number etc.
Err no.: Type: Usertext:
-121 E File `<ARG1>` has incorrect checksum!
-120 E The user management file is corrupt. Standard parameters will be set. Please change (and add) parameters in the service
menu and save them!
-119 E The contents of the file does not match the expected file structure.
-118 E The stored calibration date of the external pressure sensor inlet is invalid
-117 E The stored calibration date of the internal pressure sensor is invalid
-116 E Error when writing the calibration date of the external pressure sensor inlet
-115 E Error when reading the calibration date of the external pressure sensor inlet
-114 E Error when writing the calibration date of the internal pressure sensor
-113 E Error when reading the calibration date of the internal pressure sensor
-112 E Audit trail file is damaged. Incorrect checksum. Please archive data to create a new audit trail file.
-111 E Time exceeded when stopping the test unit
-110 E The system log file could not be stored!
-109 E The configuration file of the test unit is corrupt. The standard parameters will be set. Please change parameters in the ser-
vice menu and save them!
-108 E The configuration file of the management unit is corrupt. Standard parameters will be set. Please change parameters in the
service menu and save them!
-107 E The stored calibration protocol of the external pressure sensor is damaged. Please calibrate the external pressure sensor.
-106 E The stored calibration protocol of the internal pressure sensor is damaged. Please calibrate the internal pressure sensor.
-105 E The test program ‘<ARG1>’ could not be printed. Error: ‘<ARG2>’
-104 E The test program ‘<ARG1>’ could not be deleted.
-103 E The test program ‘<ARG1>’ could not be saved. The SD card is full.
-102 E The test program ‘<ARG1>’ could not be saved.
-101 E The test result ‘<ARG1>’ could not be printed. Error: ‘<ARG2>’
-100 E The test result could not be deleted.
-99 E The test result could not be saved to the SD card.The SD card is full.
-98 E The test result could not be saved.
-97 E The log file could not be saved to the SD card. The SD card is full.
-95 E Calibration interval of the external pressure sensor inlet invalid
-94 E Calibration interval of the internal pressure sensor invalid
-93 E Error when writing the calibration interval of the external pressure sensor inlet
-92 E Error when reading the calibration interval of the external pressure sensor inlet
-91 E Error when writing the calibration interval of the internal pressure sensor
-90 E Error when reading the calibration interval of the internal pressure sensor
-89 E Test results are damaged. The checksum is incorrect!
-88 E Cannot close touch screen file! Error: <ARG1>
-87 E Cannot open touch screen file! Error: <ARG1>
-86 E Error when setting touch screen coordinates. Error: <ARG1>
-84 E Error when stopping the test unit
-83 ... -82 E Calibration data of the internal pressure sensor invalid