72 Troubleshooting
Err no.: Type: Usertext:
-81 E Error when writing the calibration data from the external pressure sensor inlet
-80 E Error when reading the calibration data of the external pressure sensor inlet
-79 E Error while writing the calibration data from the internal pressure sensor
-78 E Error when reading the calibration data from the internal pressure sensor
-76 E New main memory cannot be allocated
-75 E Unexpected file end when reading file
-74 E Software versions of test unit, data management unit and communication server are incompatible Software version of test
unit = <ARG1> Software version of data management unit = <ARG2> Software version of xbpi2 server = <ARG3>
-73 E System error file is too large
-71 E Error when opening system error file
-70 E Error when loading test result
-67 E Program is damaged. The checksum is incorrect!
-66 E Cannot terminate communication with the test units
-65 E Cannot start communication with the test units
-64 E System partition could only be mounted in readable condition!
-63 E System partition could only be mounted in an unwritable condition!
-62 E Cannot write to “sc4user” file. Please notify your system administrator
-61 E Cannot read “sc4user” file. Please add users or notify your system administrator
-49 E Cannot open audit trail file
-48 E Cannot open error log file of the data management unit
-47 E Cannot open error log file of the test unit
-46 E Cannot open message window
-45 E Unknown test method number ‘<ARG1>’
-44 E Test unit number ‘<ARG1>’ outside defined ranged
-43 E Cannot open configuration file of test unit ‘<ARG1>’ to write.
-42 E Cannot open configuration file of test unit ‘<ARG1>’ to read.
-41 E Cannot open configuration file of processing unit to write to memory.
-40 E Cannot open configuration file of processing unit to read from memory.
0 ... 2500 E Internal error
6001 E You cannot save any more test results. Please archive the result database.
6002 W A total of ‘<ARG1>’ out a max. ‘<ARG2>’ results have been saved. Please archive the results database.
6003 E There may be more than ‘<ARG1>’ measuring points. A maximal of ‘<ARG2>’ measuring points are allowed.
6004 E Min. BP must be less than max. BP
6005 E Limits for min. BP, max. BP: 100 mbar <= min. BP, max. BP <= 8000 mbar (1.45 psi < min. BP, max. BP <=116.03 psi) mbar
and hPa values have no decimal places, psi values have up to 2 decimal places.
6006 E Limits for volume: 10 ml <= vol <= 150000 ml
6007 E The test pressure must be less than the min. BP
6008 E Limits for times: 1 min <= stability time, test time <= 99 min
6009 E Limits for test pressure: 100 mbar <= test pressure <= 8000 mbar (1.45 psi < test pressure <= 116.03psi) mbar and hPa val-
ues have no decimal places, psi values have up to 2 decimal places.
6010 E Limits for diffusion: 0.1 ml/min <= max. diffusion <= 999.9 ml/min Diffusion values can have up to 1 decimal place.
6011 E Limits for pressure drop: 1 mbar <= max. pressure drop < test pressure (0.01psi < max. pressure drop < test pressure) mbar
and hPa values have no decimal places, psi values have up to 2 decimal places.
6012 ...
6013
E Limits for test water intrusion / water flow: 0.01 ml/min <= Max. water intrusion / max. water flow <= 99.99 ml/min (0.1
ml/10min <= Max. water intrusion / max. water flow <= 999.9 ml/10min) ml/min values have up to 2 decimal places,
ml/10min values have up to 1 decimal place.
6014 E Limits for min. and max. test pressure: 100 mbar <= min., max. test pressure <= 8000 mbar (1.45 psi < min., max. test pres-
sure <= 116.03 psi) mbar and hPa values have no decimal places, psi values have up to 2 decimal places.
6015 E Limits for MPD pressure increment: 25 mbar <= MPD pressure increment <= 500 mbar (0.36 psi < MPD pressure increment
<= 7.25 psi) MPD pressure increment < 0.5*test pressure difference mbar and hPa values have no decimal places, psi values
have up to 2 decimal places.