31
System Settings (Setup Menu)
User Manual Secura
®
, Quintix
®
, Practum
®
6.2.3 Device Information
Device information
Manufacturer: Sartorius
Model: SECURA2102
Serial number: 0012345678
Version BAC: 00-50-02.xx CN:1701
Version APC: 01-70-02.xx CN:6587
The manufacturer, model, serial number, and software version of your balance are
displayed here.
6.2.4 Calibration/Adjustment
Calibration/Adjustment
isoCAL
Calibration report
With Secura
®
models, it is possible to set how the automatic calibration/adjustment
function isoCAL is started.
With Secura
®
and Quintix
®
models, saved reports about the most recent calibration
procedures, including deviations identified, can be viewed and output to a PC or
Sartorius lab printer.
isoCAL menu option (Secura
®
only):
t Select the desired start option for isoCAL.
Off: isoCAL is switched off.
Info, manual start: A note appears on the display when predefined time
intervals or temperature values are exceeded. isoCAL can be started
manually.
Automatic: isoCAL is switched on. Calibration/adjustment starts
automatically when predefined time intervals or temperature values are
exceeded.
Calibration report menu option (Secura
®
and Quintix
®
only):
t Select the desired report by date.
If there is more than one calibration report per date, use the < and > buttons to
navigate through the reports.
t Select a to print the displayed report.
A maximum of 99 entries are saved per day.
Calibration reports are deleted after 30 days.
6.2.5 Weighing
Weighing
SQmin
Application
Stability signal
Ambient conditions
Safety level
A number of basic settings for the weighing functions can be defined here. Options
are limited for verified balances. All settings given in the menu are permitted.
SQmin menu option (Secura
®
only):
If the SQmin function is set up on the balance, the function can be switched On or
Off here (see “SQmin Minimum Weighing (Secura
®
only)” on page 44).
Calibration report
2014-10-04: 2 Protocols
2014-10-01: 2 Protocols
2014-09-25: 3 Protocols
2014-09-24: 3 Protocols
2014 09 20: 4 Protocols