34
System Settings (Setup Menu)
User Manual Secura
®
, Quintix
®
, Practum
®
6.2.6 Printout
Printout
Printout
Manual print
Automatic print
You can configure the settings for printing and data output here.
Some of the settings depend on the USB port configuration (see page 36).
Under Printout:
t Define the settings for outputting logs and reports to a connected lab printer.
– Manual: The data is printed out when a is selected in the application.
– Automatic: The data is printed out automatically.
– Without stability: No stability is required for a printout.
The printout does not have any unit symbols in this case.
– With stability: The data can only be printed out when stability is reached.
– Print after weight change: Data is printed once, after a threshold has been
exceeded at stability and the weight has previously been reduced to less than
half of the threshold.
Under Manual print:
This option can be selected when Manual print has been set in the Printout menu.
t Define the settings for the Manual print.
–
ISO/GLP printout (Secura
®
and Quintix
®
only): Activate this option when an
ISO/GLP-compliant printout is required.
– Tare after print: Activate this option to automatically tare the balance after
printing.
– Manual print format: Select the information to be printed during manual
printing, such as date and value.
Value without ID: only if being transmitted to a PC (menu items: USB port:
Device/protocol: PC – xxx)
Value or Date, value or Value (N, T, G#) or Date, value (N, T, G#)
Only for Secura
®
and Quintix
®
Under Automatic print:
This option can be selected when Automatic print has been set in the Printout menu.
t Define the settings for the automatic print.
– Autom. print interval:
– Interval time:
Choose from the last three interval times or enter the desired interval time in
seconds via t and confirm with l (factory setting: 5 seconds).
– Select the Standard option to use a model-specific output rate for weight
determination.
– Autom. print format: Select the information to be printed during automatic
printing, such as date and value.
Value without ID: only if being transmitted to a PC (menu items: USB port:
Device/protocol: PC – xxx)
Valueor Date, value: only for Secura
®
and Quintix
®