39
System Settings (Setup Menu)
User Manual Secura
®
, Quintix
®
, Practum
®
6.2.11 More Settings
More settings
Set password
Reset settings
USB mass storage
Enable service mode
Additional service functions for the balance can be found here.
Set password menu option (Secura
®
only):
Secura
®
models can be password-protected to prevent unauthorized personnel from
making changes to the system settings of the balance.
When a user password is set, all settings which could change the metrological
behavior of the balance are locked.
t Select t.
t Enter a number as a password.
To make corrections to the entry:
k button: delete individual characters.
C button: delete all characters.
t Enter the password again to ensure correct entry.
t Confirm withl.
t The new password will only be applied after returning to the first level of the
menu.
To change the password, the old password must first be entered.
A new password can then be set.
To completely delete the password and use the balance without password protection,
leave the entry blank.
Contact the Sartorius Service Center if you forget the password.
Reset settings menu option:
Reset all balance settings to the factory (default) settings here.
t When the prompt appears, select Yes, reset and confirm withl. The balance is
reset and restarts.
USB mass storage menu option:
Use this function to register the balance memory as a USB removable data carrier on
a PC. This function is required by the Sartorius Service Center or when installing the
PC driver for the virtual COM port. Further information can be found in Chapter “USB
Port” on page 79.
Enable service mode menu option:
This function is used by the Sartorius Service Center and can only be accessed by
authorized Service Center personnel. The following service functions are available:
Preset value for minimum weighing (SQmin)
Enter date for next service
Reset all settings to factory settings
The Calibration/Adjustment menu also provides the following functions for
authorized Service personnel (depending on the model):
External linearization (not for use in legal metrology)
Set the preload
– Delete the preload
These functions are not available for verified and sealed balances.