7.4 Mixing (Secura
®
and Quintix
®
only)
Purpose: Use this application to weigh up to 99 components one after the other for a mixture
or formula in one container. The balance is automatically tared after each component
is weighed. The weight value of an individual component or the total weight can be
displayed as desired.
Example: You want to weigh several components of a formula in one container.
t Select the Menu key in any application.
t Select the Mixing symbol in the menu.
Max 2100 g d
=
0.01 g
0.00
Component
1
y The Mixing application appears.
Component
1
t To change settings, select the gray button.
Mixing
Print component
Off
On
y The Mixing settings window appears.
t Choose whether the individual weights of the components are printed out when
a printer is connected.
t Select
l to confirm.
t Select A to zero the balance if necessary.
t Place a container on the weighing pan.
t Select D to tare.