7.12 Peak Hold
Purpose: Use this application to calculate the maximum weight value of a sample (peak value).
The value remains on the display for five seconds after the sample has been removed
from the balance.
Example: Use for measuring release force during an experiment, or use when a load is so big
that the balance display is hidden from view during weighing.
t Select the Menu key in any application.
t Select the Peak hold symbol in the menu.
Max 2100 g d
=
0.01 g
0.00
y The Peak hold application appears.
t To define the setting for calculating the peak hold, select the gray button.
Peak hold
Apply
Without stability
At stability
y The Peak hold settings window appears.
t Choose whether the peak hold (max. value) is calculated At stability or
Without stability.
– The At stability setting is suitable for stable samples. It ensures that the
weight value fluctuations when placing or removing the sample from the
balance do not cause the calculated peak value to be incorrect.
– The Without stability setting is suitable for unstable samples.
t Select l to confirm.
t Select A to zero the balance if necessary.
t Select w.
t Place the sample on the weighing pan.