70
Weighing
User Manual Secura
®
, Quintix
®
, Practum
®
7.13 Counting
Purpose: Use this application to determine the number of parts of approximately equal weight.
The weight of a counted reference sample is calculated and then the objects with an
unknown piece count are weighed. The balance displays the number of parts and the
piece weight.
Minimize counting errors:
– Ensure that the weight of individual parts is evenly distributed.
– Larger reference piece counts lead to a higher degree of accuracy.
– Switch on Auto Reference Updating.
t Select the Menu key in any application.
t Select the Counting symbol in the menu.
Max 2100 g d
=
0.01 g
0.00
Reference
pcs
5
y The Counting application appears.
The piece count of the reference sample is displayed under Reference.
Reference
pcs
5
t To set the reference sample quantity, select the gray button.
Counting
Reference Accuracy
5
10
20
Normal
10fold
y The Counting settings window appears.
t Enter the reference piece count on the left.
Select a value or select t, enter the desired value, and confirm with l.
t Select the accuracy of the weight value during counting in the center.
y Activate “Automatic Reference Sample Updating” on the right.
This “updating” optimizes the counting accuracy. The average piece weight is
recalculated every time the reference sample is updated automatically. Since the
new parts enlarge the basis for the calculation, this increases the counting
accuracy of the reference and with it the result too.
t Select l to confirm.
t Select A to zero the balance if necessary.
t Place a container on the weighing pan.