Calibration and Adjustment
User Manual Secura
®
, Quintix
®
, Practum
®
8. Calibration and Adjustment
Background During calibration, a calibration weight is used to determine how much the displayed
value deviates from the actual measurement value. This deviation is compared against
a preset target value. The subsequent adjustment corrects this deviation or reduces
the permissible error limits.
In Secura
®
, Quintix
®
and Practum
®
balances, calibration and adjustment are combined
as one process. The balance is automatically adjusted after each calibration.
Before using a verified balance for legal metrology, calibration/adjustment
must always be carried out at the balance setup location. This can be performed
automatically or manually, or for Practum
®
models of accuracy class k using an
external weight.
When and How Often To achieve the highest accuracy possible, regularly calibrate and adjust the balance:
daily after switching on the balance,
each time the balance is leveled,
each time ambient conditions have changed (temperature, humidity, or air
pressure),
each time the balance is set up at a new location or moved in its current location.
The following describes in detail the available options for calibrating and adjusting
the balance:
Calibration/adjustment using an internal calibration weight (Secura
®
and Quintix
®
only)
Calibration/adjustment using an external calibration weight
(see “Calibration/Adjustment Using an External Calibration Weight” on
page 75)
– Automatic calibration/adjustment with isoCAL