EasyManua.ls Logo

Sartorius SIWRDCP - Preparation for Calibration and Adjustment

Sartorius SIWRDCP
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
(
1 Zero the scale
J
2 Start calibration
(e.g., after calibration prompt:
flashing
WP symbol).
c.ext.def is shown for
two seconds.
The prompt for the calibration
weight is shown
(in this example, 10 kg).
3 Place calibration weight on
the weighing platform.
The difference between
measured value and
the true mass is shown,
plus or minus sign.
Ext. calibration Printout is generated,
Nom. + 10000 g if adjustment was not
Diff. + 1 g performed and the procedure
is canceled by pressing (.
J 4 Start adjustment
(or cancel calibration/adjust-
ment by pressing ().
After adjustment,
the calibration weight value
is displayed.
-------------------- A GMP-compliant printout
24.10.2006 10:15 is generated
Typ SIWR
Ser.no. 12345678
Vers. 1.0103.11.2
BVers. 01-26-02
--------------------
Ext. calibration
Nom. + 100000 g
Diff. + 10 g
Ext. adjustment
Diff. + 0.0 g
--------------------
24.10.2006 10:15
Name:
24.10.2006 10:15
45
Example:
External calibration and manual adjustment with default weights (with factory settings for weighing parameters)