Control Drawing
2012-07-25
Dr. D. Klausgrete
Drawing
number
Title
Revision 03
1
NOTES
1) USA: The installation must be in accordance with the National Electrical Code
®
, NFPA 70, Article 504 and ANSI /
ISA-RP 12.6.
Canada: The installation must be in accordance with the Canadian Electrical Code Section 18.
2) The Apparatus must not be connected to any device that uses or generates in excess of 250Vrms or DC.
3) USA: The Apparatus must be connected to a suitable ground electrode per National Electrical Code
®
, NFPA 70,
Article 504. The resistance of the ground pad must be less than 1 ohm.
Canada: The Apparatus must be connected to a suitable ground electrode per
Canadian Electrical Code ®, Part 1. The resistance of the ground pad must be less than 1 ohm.
4) Connection by non interchangeable cable Type LiYC-Y-CY 4x0.5.
5) WARNING: SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
AVERTISSEMENT: LA SUBSTITUTION DE COMPOSANTS PEUT COMPROMETTRE LA
SÉCURITÉ INTRINSÈQUE.
6) Series FCT01-X.. control drawing 65638-000-07-A4
Series FC/IS/FCA/FCB……-.X… control drawing 33956-000-07-A4
Series WZ……-X…... control drawing 35520-001-07-A4
Series CIXS3-. / CW3XS.U-……-….. control drawing 65607-000-07-A4
For Installation in the USA:
Hazardous (Classified) Location
XP Class I, II, III, Division 1, Groups A,B,C,D,E,F,G
AIS for Class I, II, III, Division 1, Groups A,B,C,D,E,F,G
SARTORIUS
Model YPS02-XUR
SARTORIUS
IS Equipment
4)
3)
2)
AC Supply
Hazardous (Classified) Location
see Control Drawing of the IS equipment
4)
For Installation in Canada:
Hazardous Location
XP Class I, II, III, Division 1, Groups B,C,D,E,F,G
[Ex ia] ASSOCIATED EQUIPMENT Class I, II, III,
Division 1, Groups A,B,C,D,E,F,G
SARTORIUS
IS Equipment
Hazardous Location
see Control Drawing of the IS equipment
SARTORIUS
Model YPS02-XKR
3)
2)
AC Supply
6)
6)
DiV 2: Ta 50°C