Laser Beam Couplers Series 60SMS
Adjustment_60SMS 10.12.2020 © Schäfter+Kirchhoff GmbH
Schäfter+Kirchhoff GmbH Kieler Str. 212 22525 Hamburg, Germany +49 40 85 39 97-0 info@sukhamburg.de www.sukhamburg.com
7
Figure 15:
Laterally shift the adapter in order to
maximize the radiation that is trans-
mitted by the aperture.
Only a coarse alignment is necessary, which can be done by hand (Figure 15), as the positioning accuracy
must only be a fraction (7 - 10 %) of the collimated beam diameter.
Figure 16:
Finally, fix the adapter and remove the
aperture.