SEAWARD Electronic Rigel 266 Plus - User Manual

SEAWARD Electronic Rigel 266 Plus
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Rigel 266 Plus
Compact Electrical Medical Safety Analyser
Instruction Manual
350A555 Issue 1.0
February 2006
© 2006 Seaward Electronic Ltd.
Issue 1.0

Table of Contents

Questions and Answers

  • C
    caitlintateAug 3, 2025
    What to do if SEAWARD Electronic Measuring Instruments shows LOAD TOO GREAT Err5?
    • C
      Crystal MyersAug 3, 2025
      If your SEAWARD Electronic Measuring Instruments displays 'LOAD TOO GREAT Err5', it indicates that the appliance under test might be drawing more than 18A. Use your discretion; if you're unsure, don't proceed with the test and seek professional advice.
  • C
    christopherknightAug 13, 2025
    How to fix HOT Err7 on SEAWARD Electronic Measuring Instruments?
    • V
      veronica11Aug 13, 2025
      If your SEAWARD Electronic Measuring Instruments displays a 'HOT Err7' message, allow the analyzer to cool down before pressing Enter to continue testing.
  • S
    Stephen JonesAug 23, 2025
    Why does my SEAWARD Electronic Measuring Instruments show INTERNAL FAULT Err4?
    • M
      Mark MyersAug 23, 2025
      The 'INTERNAL FAULT Err4' error on your SEAWARD Electronic Measuring Instruments could be due to the analyser’s internal over temperature switch activating, especially if prolonged earth bond tests have been carried out previously. If the problem persists, the Rigel 266 plus is faulty and requires repair. Return the unit to Rigel Medical for correction of the fault.

Summary

Before Starting

Safety Information

General Safety Precautions

Warnings and precautions for operating the analyser safely.

Chapter 2 Using the Analyser

Connecting the Analyser

Powering and Connecting Procedures

Instructions for connecting the analyser to a mains supply and initial power-on.

Performing a Test

Test Modes and Execution

Details on test modes, execution, aborting, and printing results.

Equipment Type Selection

Insulation Tests

Insulation Test Setup and Procedure

How to connect appliances and perform insulation tests.

Powered Tests

Load Test Operation

Procedure for conducting the Load Test.

Earth Leakage Test

Earth Leakage Test Procedure

How to perform Earth Leakage tests, including duration and conditions.

Enclosure Leakage Test

Enclosure Leakage Test Procedure

How to perform Enclosure Leakage tests, including settings and conditions.

Patient Leakage - Applied Part To Earth

Patient Leakage Test Setup and Safety

Safety warnings and connection instructions for Patient Leakage tests.

Patient Leakage - F-Type

Patient Leakage - Mains on SIP;SOP

Patient Auxiliary Current

IEC Lead Polarity Test

Chapter 3 Manual Mode

Chapter 4 Tips & Troubleshooting

Power On Self-Test Issues

Troubleshooting common error messages during power-on self-tests.

Temperature Monitoring

Interfacing Options

Serial Port Configuration

Chapter 5 Maintaining the Analyser

Service and Calibration

Chapter 7 Specifications

Mechanical and Environmental Specs

SEAWARD Electronic Rigel 266 Plus Specifications

General IconGeneral
BrandSEAWARD Electronic
ModelRigel 266 Plus
CategoryMeasuring Instruments
LanguageEnglish