USER
RESPONSIBILITY
CAUTION
Federal
law
restricts
this
device
to
sale
by
or
on
the
order
of
a
physician.
|
The
SECHRIST
Air-Oxygen
Mixer
is
a
restricted
medical
device.
The
SECHRIST
Mixer
will
perform
in
conform-
ance
with
the
specifications
and
description
contained
in
this
Calibration
and
Service
Manual
and
other
accom-
panying
labeling
when
assembled,
operated,
maintained
and
repaired
in
accordance
with
the
instructions
contained
in
this
manual
and
other
acccompanying
documentation.
Do
not
attempt
to
operate
this
Mixer
before
reading
and
thoroughly
understanding
these
instructions.
The
Mixer
must
be
checked
periodically
as
specified
in
this
manual
(see
Routine
Maintenance
Section).
A
defective
device
should
not
be
used.
Should
repair
become
necessary,
this
Calibration
and
Service
Manual
is
available
to
qualified
hospital
technicians.
If
the
hospital
does
not
have
the
capability
to
calibrate,
service
and
repair
this
equipment.
please
contact
the
authorized
SECHRIST
distributor
who
effected
delivery
of
this
device.
If
a
suitable
response
is
not
received
within
a
reasonable
time,
communicate
directly
with
our
Home
Office
in
Anaheim,
California.
This
product
should
be
repaired
only
in
accordance
with
written
instructions
from
SECHRIST,
and
should
not
be
modified
in
any
way,
except
with
prior
written
approval
from
SECHRIST.
WARNING
Whenever
a
patient
is
attached
to
respiratory
care
equipment,
constant
attendance
is
required
by
qualified
personnel.
The
use
of
an
alarm
or
monitoring
system
does
not
give
absolute
assurance
of
warning
for
every
form
of
malfunction
that
may
occur
with
the
system.
In
addition,
some
problems
may
require
immediate
attention.
SUMMARY
OF
WARNINGS
AND
CAUTIONS
The
SECHRIST
Air-Oxygen
Mixer
is
a
restricted
medical
device
intended
for
use
only
by
qualified
personnel
under
the
direction
of
a
qualified
physician.
Personnel
operating
this
equipment
are
responsible
for
reading
and
thoroughly
understanding
all
product
documentation
provided.
Statements
throughout
the
product
documentation
provided
have
special
significance
as
follows:
KEY
TO
WARNINGS,
CAUTIONS
AND
NOTES:
WARNING
Means
there
is
a
possibility
of
personal
injury
to
the
operator
or
patient.
CAUTION
Indicates
there
is
a
possbility
of
damage
to
the
Mixer
or
other
equipment
attached
to
it.
NOTE:
Notes
are
used
to
call
attention
to
statements
pertaining
to
more
efficient
or
convenient
operation
or
service
of
the
equipment.
SUMMARY
OF
ALL
WARNINGS
AND
CAUTIONS
CONTAINED
IN
THIS
DOCUMENT:
WARNING
Whenever
a
patient
is
attached
to
respiratory
care
equipment,
constant
attendance
is
required
by
qualified
personnel.
The
use
of
an
alarm
or
monitoring
system
does
not
give
absolute
assurance
of
warning
for
every
form
of
malfunction
that
may
occur
with
the
system.
In
addition,
some
problems
may
require
immediate
attention.
WARNING
This
precision
gas
mixing
device
is
intended
for
use
with
Medical
Breathing
Air
and
Medical
Breathing
Oxygen
only.
Do
not
modify
inlets
to
accommodate
other
breathing
gases.
especially
anesthetic
gases.
CAUTION
Federal
law
restricts
this
device
to
sale
by
or
on
the
order
of
a
physician.
CAUTION
Grasp
nipple
at
base
next
to
the
proportioning
valve
block
noton
the
end
or
the
nipple
may
be
damaged.
CAUTION
If
resistance
is
met
before
stop
pin
at
1.0
position
is
reached,
do
not
force
the
knob.
Re-adjust
rear seat
slightly
just
enough
to
allow
free
range
of
movement
to
stop
pin
position.
Recheck
the
calibration
as
described
above.
CAUTION
Alarm
bleed
set
screw
(P
/
N
3568)
is
calibrated
at
factory
and
should
not
require
calibration.