21
-
6-2. Setting humidity recording range
①
Before setting, make sure that the DIP switches No. 3
and No. 4 are turned ON.
②
Press the
【
FUNCTION
】
key six times (Seven times
when the SD card is inserted and the [SD Card Re-
cord] switch is set to OFF) while the normal mea-
surement screen is displayed.
『
MODE
』
and
『
r2
』
will
appear and the present central humidity of the chart
will blink on the left side.
『
- -
』
will blink instead if the
recording range has not been changed from the initial
setting. (0 to 100% RH)
③ Change the humidity to be recorded in the center of
the chart paper with the
【▲】
or
【▼】
key, then press
the
【
ENTER
】
key. The central humidity will stop blink-
ing to show the setting has been completed.
④ Then, the present recording range will blink on the
right side of the screen.
⑤ Set the recording range 10 (+ / -10%) or 25 (+ / -25%)
from the central humidity with the
【▲】
or
【▼】
key,
then press the
【
ENTER
】
key. The recording range
will stop blinking to show that setting has been com-
pleted.
Reference
If this setting is not required, press the
【
FUNCTION
】
key once to return to the normal display.
FUNC-
7-1. Function overview
Set the alarm for the abnormal value to inform you when
measured, with the
『
H
』
or
『
L
』
indicator blinking in the
LCD.
Reference
If the alarm has been set for 30.0 °C / 50.0 °F (50.0 %)
as the upper limit and the measured value is 30.0 °C /
50.0 °F (50.0 %) or higher, the
『
H
』
indicator will appear
and start blinking. If the current value becomes 29.5 °C
/ 49.5 °F (49.0 %) or lower,
『
H
』
will go out.
Also, if the same value has been set as the lower limit
and the measured value is 30.0 °C / 50.0 °F (50.0 %)
or lower, the
『
L
』
indicator will appear and start blinking.
If the current value becomes 30.5 °C / 50.5 °F (51.0 %)
or higher,
『
L
』
will go out. (The hysteresis is 0.5°C or 0.5
°F in temperature and is 1.0% RH in humidity.)