TB-6123 Revision 10/06
Page 11 of 12
5. Select the “10M safe” or “35M safe” setting, whichever one is appropriate, on the
62080/62100/62102 and repeat the test. The results of the test should be “pass” for the
wrist strap.
6. Select the “10M high” or “35M high” setting, whichever one is appropriate, on the
62080/62100/62102 and repeat the test. The results of the test should be “fail high” for the
wrist strap.
Footwear operation test.
1. Insert the 62080/62100/62102 test plug into the foot plate cable connector on the
62100/62102.
2. Select the appropriate “fail low” range on the rotary switch of the 62080/62100/62102.
3. Press and hold the touch plate of the 62100/62102 until the test is completed. The results of
the test should be “fail low” on both feet.
4. Select the appropriate “safe low” setting on the 62080/62100/62102 and repeat the test. The
results of the test should be “pass” on both feet.
5. Select the appropriate “safe high” setting on the 62080/62100/62102 and repeat the test.
The results of the test should be “pass” on both feet.
6. Select the appropriate “fail high” setting on the 62080/62100/62102 and repeat the test. The
results of the test should be “fail high” on both feet.