HOW THE INSPECTOR WORKS:
An electrochemical sensor behind the sensor inlet shown in the image above will create an electrical signal when
carbon monoxide (CO) is sensed. The gas concentration is displayed in parts per million (PPM).
HOW TO PERFORM A SELF CALIBRATION:
For reasonable accuracy a frequent calibration is recommended at least once every 6 months. Sensorcon will
perform this procedure for you or you can do it yourself if you have the necessary 50ppm calibration gas.
Please follow these steps to self calibrate your Inspector:
1. Hold both buttons for 5 seconds to enter calibration mode. Press the MUTE button at any time to exit.
2. In a clean air environment then press the MAX button to begin the ZERO calibration. Wait for countdown.
3. When zero is complete, use 1/4” OD, 1/8” ID Tygon tubing attached to the calibration gas cylinder and
sensor inlet to flow calibration gas at 0.5L/min, then press the MAX button to start calibration process.
4. The Inspector will then automatically calibrate itself and notify you if a calibration is successful.
Please feel free to contact us if you have any further questions or are in need of technical assistance.
Please Note: The Sensorcon Inspector is a professional tool. It does not require very much maintenance with the exception of keeping it clean.
This is especially important in the area of the sensor inlet. If it gets dirty then simply wipe it clean with a damp cloth or with a damp cotton swab.
Sensor Inlet
Hold down both buttons to begin.
Connect the 50ppm calibration gas.
Wait for the countdown to complete.
150 Airport Drive Buffalo, NY 14225 Office: (716) 566-2728 Fax: (716) 276-3047 sales@sensorcon.com www.sensorcon.com
Page 3
Question and Answer IconNeed help?

Do you have a question about the Sensorcon CO Inspector and is the answer not in the manual?

Sensorcon CO Inspector Specifications

General IconGeneral
BrandSensorcon
ModelCO Inspector
CategoryMeasuring Instruments
LanguageEnglish

Summary

Product Warranty and Operating Conditions

Warranty Details

Provides information on the product's one-year warranty and the importance of regular calibration.

Normal Operating Conditions

Specifies the optimal temperature and humidity ranges for the device's normal operation.

Practical Considerations for Use

Discusses the impact of humidity and temperature extremes on sensor performance and potential damage.

Important Notes on Carbon Monoxide

Explains what Carbon Monoxide is and the role of the Inspector in detecting this toxic gas.

Device Layout and Display

Inspector Layout Overview

Details the physical components of the Inspector device, including the sensor inlet and buttons.

LCD Display Layout Explained

Describes the various indicators and icons displayed on the device's LCD screen during operation.

Device Operation and Calibration

How the Inspector Works

Explains the electrochemical sensor principle behind detecting carbon monoxide concentration.

Performing a Self Calibration

Provides step-by-step instructions for users to calibrate the Inspector device themselves.

Product Model Comparison

Model Feature Comparison Chart

Compares features and specifications across Standard, Industrial, and Industrial PRO models.

Company Commitment and Purchase

Highlights Sensorcon's dedication to quality and information on purchasing products online.