14
HEPA CO
2
FILTER
The CO
2
Incubator design features a HEPA filtration system with a patented copper
housing. (Patent Number 6,333,004) As the chamber air is drawn through the filter
system, airborne microbes and isolated particulates are not only removed from the air,
but destroyed as well. It is through copper oxidation that the contaminants are actually
killed. The HEPA filter is easily replaced without the need of any tools and has an
efficiency rating of 99.97% at 0.3 microns.
Install the HEPA Filter onto the duct prior to installing the duct in the unit.
WARNING: Ensure that when installing the duct to ensure that the duct is not
touching the C02 sensor head as this may cause erroneous readings.2800517.
Section
8
The Hepa filter attaches to a fixed
coupling ring with a direct “nap-in”
action. Push straight up to the
coupling ring to connect. Pull the
filter straight down to disengage.
Replacement Hepa filters is
available from Sheldon Mfg., Inc.
Part Number 2800517.