Shimadzu AA-7000 Series - Specialized Accessories and Systems; Dual Atomizer System & Lamps; High-Sensitivity Analysis Accessories

Shimadzu AA-7000 Series
28 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Optional Accessories / Related Products
For AA-7000G
For AA-7000F
206-52135-91
206-83755-91
Dimensions: 170 W × 110 D × 110 H mm
I.D.: ø25 × 170 mm
Heater: PYROMAX DS
Furnace
Controller
Operational temperature range: Ambient to 1100 °C
Temperature regulation range: ±2.5 °C
Load capacity: 1000 W max.
Power requirements: 100 VAC, 400 VA, 50/60 Hz
For Higher Sensitivity with the Hydride Generation Method
* Use a stepdown transformer in regions that do not have a 100 V power supply.
SARF-16C atomic muffle furnace
(Electronic Cell Heater)
208-97249
ComputerHVG-1 hydride vapor generator
Controller
Temperature sensor
Cell heater
Atomic absorption
spectrophotometer
Te
Cell
he
Mounting adaptor
For High-Sensitivity As, Se, Sb Analysis
HVG-1 hydride vapor generator 206-17143-XX
Environmental standards prescribe the hydride generation method as one
method of As, Se, and Sb analysis.
1. Used with the AA-7000 Series, HVG-1 permits the rapid and accurate
quantitation of elements such as As, Se, Sb and Hg at several-ppb levels.
2. Used with an ASC-7000 autosampler, it permits the automated serial
analysis of up to 60 samples.
* Nozzle ASSY, HVG (P/N: 206-67563) is required to use HVG-1 with an ASC-7000 Series
instrument. Order separately.
1. This dedicated furnace permits higher sensitivity measurements using the
hydride vapor generator than the flame heating method.
2. The temperature controller provides optimal control of the quartz cell
temperature.
3. Good temperature reproducibility. Extends cell life by preventing
damage due to overheating.
* The atomic muffler cannot be used when an AAC-7000 is installed on an AA-7000F/AAC
or AA-7000F.
Hydride vapor generator, absorption cell (P/N: 206-77607)
Reagent bottles (P/N: 206-58792-40/-42), gas hose, drain tube, etc.
Measurement method: continuous flow
Sample consumption: 0 to 7 mL/min, variable
Reagent consumption: 0 to 2.5 mL/min, variable
Atomizer: Heated absorption cell (heated by Air-C
2
H
2
flame in standard system)
Carrier gas: Ar pressure: 0.32 MPa, consumption: 70 mL/min
Power requirements: 100, 120, 220, 230, 240 VAC, 35 VA, 50/60 Hz
Dimensions: 340 W × 220 D × 200 H mm
Weight: Approx. 9 kg
I
Specifications
I
Standard Major Items
Required to add GFA-7000A to AA-7000F to conduct furnace analysis.
Permits automatic burner position setting for flame analysis with AA-7000F.
AAC-7000 auto atomizer changer 206-77701-41
Dual Atomizer System
For SR method
L-233 series
L-2433 series
200-38422-XX
200-38456-XX
Hollow Cathode Lamps
Used for analog output to a pen recorder, etc.
One cable required per channel (atomic absorption/energy signal, background signal).
Analog output cable 206-77707-91
Other Accessories
Part Name P/N Remarks
Part Name P/N Remarks
Part Name P/N Remarks
Part Name P/N Remarks
Part Name P/N Remarks
I
Specifications
22