The PC, monitor and printer are not included in the standard configuration. Prepare them separately.
.
The ASC-7000, ASK-7000 and GFA-7000A are optional accessories.
.
The separately available micro sampling kit is required for the flame micro sampling method.
.
Microsoft Windows 7 Professional / Vista Business and XP Professional are registered trademarks of Microsoft Corp., U.S.A.
.
The AA-7000 Series does not conform to IEC60601 –– Medical Electrical Equipment.
Flame Continuous Method
Required Software Operation Environment
Furnace Method / Flame Micro Sampling Method
Autosampler
Common Specifications
Functions Zero-point detection, auto diagnosis, auto rinse, random access
Sample volume Sample vials: 15 mL, reagent vials: 53 mL
Solvent rinse aspiration method
Sampler: 340 W × 280 D × 270 H mm, 8 kg
Controller (including stand): 260 W × 200 D × 320 H mm, 7 kg
Nozzle rinse
Dimensions and weight
Functions Zero-point detection, auto diagnosis, auto rinse, auto mixing, random access
Sample volume Sample vials: 2 mL, reagent vials: 20 mL
Solvent rinse discharge method
Sampler (including sliding parts): 340 W × 280 D × 400 H mm, 11 kg
Controller: 235 W × 110 D × 240 H mm (built into GFA-7000A), 6 kg
Nozzle rinse
Dilution function, reagent addition function
Sampling functions
250 μL
Syringe
2 to 90 μL
Injection volume
1%RSD (20 μL)
Repeatability
Rinse port: Less than 0.0001
Mixing port: Less than 0.00001
Carryover
Solvent rinse discharge method, solvent discharge and rinse with next sample
Mixing port rinse
Performed in mixing port. Maximum mixture volume: 0.6 mL
Mixing function
Up to 4 solutions
Number of solutions possible for mixing
.
Calibration curve method: 5 solutions max. (sample + 4 types of reagents)
.
Standard addition method: 6 solutions max. (sample + standard solution + 4 types of reagents)
Maximum number of
mixing reagents added
For measurement result on unknown samples:
.
If extrapolation of calibration curve is possible: automatic calculation of dilution rate and dilution to bring concentration within calibration curve range
.
If extrapolation of calibration curve is not possible: dilution rate fixed at 10x
Auto dilution / re-analysis
Dimensions and weight
Intel
®
Celeron 420 (1.60 GHz or higher)
To use GFA-TV, Intel
®
Pentium DualCore E2180 (2 GHz) or higher is required.
1 GB or higher (Vista) or 512 MB or higher (XP)
To use GFA-TV, 1 GB or higher (GFA-TV)
XGA (1024 × 768 dots) or higher
One CD-ROM drive (for installing a software)
Requires 60 MB min. hard disk space for AA installation.
One serial port (for AA control)
USB 2.0 port (for GFA-TV)
CPU
Microsoft Windows 7 Professional (32bit) / Vista Business / XP Professional
Operating system
RAM
Monitor
Storage device
I/O port
Monitor, keyboard, mouse, printerPeripheral device
Shimadzu recommends the PC and monitor in the PC Set for AA-7000 Series with WizAArd Software Pre-Installed.
Purchase a printer separately. The personal computer and monitor specifications are as follows.
RS-232C communication control from AA unitControl
Maximum reagent / sample positions
Rinse water bottle
Power requirements
Reagents: 8 positions Samples: 60 positions (random access available with reagents or samples)
2 L
Choose from 100, 120, 220, or 230 VAC, 50 VA, 50/60 Hz