Message Display Probable Cause(s) Countermeasure See:
ERR H
(Hardware error)
z
There is a fault in the hardware,
such as the temperature sensor
or internal weight mechanism
(TW series only).
z
There is an error in the internal
system data.
z
Disconnect the AC adaptor and
turn the power back ON. If the
same message is still displayed,
contact your Shimadzu
representative.
Page 29
ERR C
(Span calibration error)
z
The balance has a large drift of
the zero point or sensitivity.
z
A container is placed on the pan.
z
The pan is displaced.
z
The wrong weight has been
placed on the pan.
z
Press to return to the
weighing mode. Execute
span calibration again in an
appropriate condition.
Page 53
CAL D
z
Display is very instable.
z
Press to return to the
weighing mode and execute
calibration again when there is
no wind or vibration. If CAL D
is still displayed, contact your
Shimadzu representative.
Page 50
ERR N
(Numerical value entry
error)
z
Either a mistake has been made
when entering the value or the
value is not appropriate.
z
After the error is displayed, the
balance returns to the status
immediately before the error
occurred. Enter the correct
numerical value.
Page 45
ERR W
(Operation error)
z
The operation used is wrong.
z
After the error is displayed, the
balance returns to the status
immediately before the error
occurred. At this point, follow
the correct operation.
-
COM ERR
(External input error)
z
An unrecognizable command
code has been received.
z
After the error is displayed, the
balance returns to the status
immediately before the error
occurred. At this point, set the
correct command code.
Page 124
- OL
OL
(overload)
z
Either the pan or the pan
supporter is displaced.
z
Set the pan or pan supporter
correctly on the balance.
Page 27
z
The weighing capacity has been
exceeded.
z
Use the balance within its
weighing capacity.
Page 146
ABORT
(Operation aborted)
z
The calibration or standard
value setting operation has
been aborted.
z
After this is displayed, the
balance returns to the operable
state.
-
WAIT
(Waiting for permission for
the operation)
z
This message is displayed in
order to avoid unnecessary key
operations.
z
After this is displayed, the
balance returns to the operable
state.
-
BUSY
(Load detected)
z
There was something placed
on the pan when calibration
was started.
z
Take the item off the pan.
The message will be cleared
automatically and you will be
able to continue calibration.
Page 52
PLS.CAL
(Load detected)
z
There was something placed
on the pan when PSC was
started.
z
Execute span calibration with
nothing on the pan.
Page 56