16 Special Functions
16.7 Service and Maintenance
262
GC-2010 Plus
16.7.4 INSTALLATION (PIPING)
16.7.4.1 Screen description
Select “7. SERVICE/MAINTENANCE” from the [FUNC] key screen, and then select “2.
INSTALLATION (PIPING)”, to open the Installation (Piping) Screen shown in Fig.
16.7.4.
In the carrier gas, flow controller fields, specify where the temperature control port of the
injection unit is connected.
In the detector gas flow controller fields, specify the slot number of flow control unit and
the temperature control port of the detector unit.
16.7.4.2 Parameter list
Q Carrier gas flow controller settings
The names CAR1, CAR2 and CAR3 are automatically assigned in ascending order of the
slot No. for each installed AFC.
Specify the flow controller carrier gas settings for each CAR.
UNIT TYPE
For, display only.
When an AFC is installed, this is automatically displayed.
If a manual flow controller is installed, specify the installation Slot No. (See below.)
“SPLITTER” is automatically displayed for the Unit Type.
SLOT NO.
Selection: NON/SLOT1/ SLOT3/SLOT5/ SLOT6, Default: NON
This item can be set only when a manual flow controller is installed.
Select the Slot No. where the manual flow controller is installed.
Slot No. which have already been set cannot be selected.
The slot No. of an installed AFC is automatically recognized and displayed.
Fig. 16.7.4   Installation status setup screen Fig. 16.7.5   Installation status setup screen
      (continued)
Detector gas
flow controller
settings
Carrier gas
flow controller
settings