19
Day 3
8. S/N Measurements
Measure S/N.(EI MODE)
SAMPLE OFN 1pg/ul (isookutane solvent) 1ul
Measurement Conditions
INJ 250
M
C
SPLITLESS mode (sampling time 1.0 minute)
COL: 50
M
C (1min) - 40
M
C / min - 200
M
C (0min) - 15
M
C / min - 280
M
C
I/F: 250
M
C
IS: 200
M
C
He: during injection 250kPa, 1 minute
(HIGH-PRESSUREINJECTION mode)
during analysis 120Kpa
Total flow: 50mL/minute
MS: SCAN mode (mass range: m/z200 to 300)
interval 0.5s
detector voltage relative value: +0.4 (absolute value: 1.4 to 1.6kV)
measurement time: 3.0 to 10.0 minute (solvent elution time: 3 minutes)