Chapter 5 Options
Mobile DaRt Operation Manual
5-8
2. The reset button of the dose area meter is pushed. The display of
the display part becomes 0, and the measurement preparation of the
dose area product is completed.
NOTE
If you measure the amount of the dose area product, push the reset
button. Push the reset button can be considered to be the START
signal for the measurement of dose area product. Pushed the reset
button is recommended to be called up before any measurement.
3. X-ray is irradiation.
4. The dose area product display in the display part of the dose area
meter is confirmed.
Cleaning
Cleaning the measuring chamber with a cloth soaked in water with a
mild washing-up liquid added is possible.
Do not use scouring or solvent-containing cleansing agents (e.g. benzine,
alcohol, stain-remover) or organic solvents for cleaning the measuring
chamber.
Using such agents may result in damage to the conductive layers on the
electrodes or in reducing the transparency of the ionization chamber.
NOTE
While cleaning, make sure no moisture gets into the ionization
chamber. The chamber must be dry before starting the system
again..
NOTE
Clean the keypad by wiping it with cloth soaked in a disinfectant.

Table of Contents

Other manuals for Shimadzu MUX-100D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu MUX-100D and is the answer not in the manual?

Shimadzu MUX-100D Specifications

General IconGeneral
BrandShimadzu
ModelMUX-100D
CategoryMedical Equipment
LanguageEnglish

Related product manuals