SPECIFICATIONS 7
7.2 Technical Data
Measuring Parameters
Measured variable Transmission, opacity, relative opacity, extinction, scattered light intensity, dust
concentration
Measuring range (freely
adjustable)
Min. Max.
● Transmission
100 ... 90% 100 ... 0%
● Opacity
0 ... 10% 0 ... 100%
● Relative opacity
0 ... 10% 0 ... 100%
● Extinction
0 ... 0,045 0 ... 2
● Dust concentration
Scattered light measurement
0 - 5 mg/m³
0 - 200 mg/m³
Transmission measurement (depending on active measuring path and particle
characteristics; see special graphics)
0 - 200 mg/m³
0 … 10,000 mg/m³
Measurement uncertainty
1)
± 2%
Response time 1 ... 600 s; freely selectable
Measuring Conditions
Distance flange - flange
2)
with reflector/scattered light receiver DHC-R0
with reflector/scattered light receiver DHC-R1
0.5 ... 3 m
2.5 ... 8 m
Effective active measuring
path (scattered light
measurement)
3)
with reflector/scattered light receiver DHC-R0
with reflector/scattered light receiver DHC-R1
Approx. 0,1 m
Approx. 0,3 m
Swivel error
4)
0.2% transmission
Gas temperature (above dew
point)
-40 ... 300 °C Higher on request
Sample gas pressure -50 hPa ... +2 hPa
-50 hPa... +30 hPa
MCU-P control unit
Optional external purge air unit
Ambient temperature -40 ... +60°C
-40 ... +45°C
Sender/receiver unit, reflector/scattered light receiver,
MCU-N control unit
MCU-P control unit, intake temperature for purge air
Function Check
Automatic self-test Linearity, drift, aging, contamination
Contamination limit value: From 30% warning: From 40% failure
Manual linearity check Using a reference filter
Output Signals
Analog output 3 outputs 0/2/4 ... 22 mA, max. load 500 W (standard output max. 750 W);
resolution 10 bits; electrically isolated
2 further analog inputs when using an Analog Input module (option, see “MCU control unit”,
page 23)
Relay outputs 5 potential-free outputs (changeover contacts) for status signal; load 48 V, 1 A
Input signals
Analog input 2 inputs 0...20 mA (standard; without electric isolation); resolution 10 bits
2 further analog inputs when using an Analog Input module (option, see “MCU control unit”,
page 23)
Digital input 4 inputs to connect potential-free contacts (e.g., for external maintenance switch, triggering
function check);
Communication Interfaces
USB 1.1, RS 232 (on
terminals)
For measured value inquiries and software updates per PC/laptop using the operating
program
RS485 For connection of sender/receiver unit
Optional Interface module To communicate with the Host PC, optional for Profibus, Ethernet
1): In temperature range - 20 °C ... +50 °C
2): Upper limits only with distortion-free fitting
3): see “Measuring principle with transmission and scattered light measurement with forward dispersion”, page 11
4): With swivel angle ± 0.3 °; total swivel range ±1 °