8011952/YWL2/3-0/2016-08| SICKOPERATING INSTRUCTIONS | DUSTHUNTER C200
Subject to change without notice
2 PRODUCT DESCRIPTION
2 I/O module
Apart from the standard analog output, the DUSTHUNTER C200 has an Analog module with
two outputs 0/4 ... 22 mA (max. load 500 Ω) integrated to output further measured vari-
ables. The module is plugged onto a module carrier connected to the processor board with
a special line.
Options
1 1x analog input module with two inputs 0/4 ... 22 mA (see “Options for MCU control
unit”, page 123) to read-in values from external sensors (gas temperature, internal duct
pressure, moisture, O
2
) to calculate the dust concentration in standard state.
An additional module carrier docked to the existing ones is required for this option.
2 Module to pass on measured values, system status and service information to higher
level control systems, optionally for Profibus DP V0, Modbus TCP or Ethernet (type 1 or
type 2), to plug onto a hat rail (see “Options for MCU control unit”, page 123).
A corresponding line serves to connect the module to the processor board.
3 MCU remote control unit
The MCU remote control unit has identical functions to the MCU display near the device,
however, it can be installed further away.
Operating function the same as the MCU display
Distance to the device:
- MCU remote control unit without separate power supply unit: Max. 100 m
- MCU remote control unit with own power supply unit: Max. 1000 m
The MCU and the MCU remote control unit are interlocked (it is not possible to
operate both MCUs at the same time).
Profibus DP-V0 for transfer via RS485 according to DIN 19245 Part 3 as well as
IEC 61158.