3 ASSEMBLY AND INSTALLATION
3.2.2 Fitting the MCU control unit
Fit the MCU control unit in a protected location that is easily accessible (see “MCU
assembly dimensions”, page 38). Observe the following points during fitting:
Maintain the ambient temperature according to the Technical Data; take possible
radiant heat into consideration (shield when necessary).
Protect against direct sunlight.
Whenever possible, choose an assembly location with minimum vibrations; dampen any
vibrations when necessary.
Provide sufficient clearance for lines and opening the door.
Assembly dimensions
Fig. 20: MCU assembly dimensions
c
f
e
> 250
d
b
M8
Clearance to open door
Clearance for cables
a
Dimen-
sion
Control unit type
MCU-N MCU-P
a 160 260
b 320 420
c 210 300
d 340 440
e 125 220
f > 350 > 540
MCU-N:
Control unit without purge air supply
MCU-P:
Control unit with purge air supply
(see “MCU control unit”, page 23)