4 START-UP AND PARAMETER SETTINGS
4.3 Sender/receiver unit and reflector/scattered light receiver
After completion of the tasks described above, the sender/receiver unit and the reflector/
scattered light receiver must be removed from the adjusting stands or zero tube and taken
to the measuring location.
4.3.1 Connecting the scattered light receiver
▸ Check whether the purge air supply is available (the flow direction must be correct and
the purge air hoses fitted tight on the connections).
▸ For purge air supply via the MCU-P control unit or external purge air supply, push the
purge air hose on the connections of the sender/receiver unit and the reflector and
secure with hose clamps.
4.3.2 Fitting and connecting on the duct
▸ Attach the sender/receiver unit and reflector/scattered light receiver to the flange with
tube and secure with the associated assembly kit (see “Assembly parts”, page 122),
tighten the self-locking nuts as tight as possible.
Fig. 42: Fitting the sender/receiver unit / reflector on the duct
C
AB
1 pair
Sender/receiver unit
Self-locking nut
Spherical washer
Sealing tape
Cup springs (4 pairs); only with fastening set for sender/receiver unit
A
Horizontal alignment
B
Fixing point
C
Vertical alignment