8011952/YWL2/3-0/2016-08| SICK OPERATING INSTRUCTIONS | DUSTHUNTER C200
Subject to change without notice
START-UP AND PARAMETER SETTINGS 4
4.4.2 Factory settings
The steps required to modify these settings are described in the following Sections. For this
purpose, the devices must be connected in SOPAS ET (see “Connection to the device via
USB line”, page 55), the Level 1 password set and the “Maintenance” mode set.
Parameter Value
Function check Every 8 h; output of control values (every
90 s) on standard analog output
Analog output (AO)
[mA]
Live zero (LZ) 4
Upper measuring range value
(MBE)
20
Current during maintenance 0.5
Current by malfunction 21 (optional 1)
Response time 60 s for all measured variables
Measured variable
Output on AO Value at LZ Value at MBE
Opacity [%]
1 0 100
Dust concentration a.c. ext [mg/m³]
2 0 200
Dust concentration a.c. SI [mg/m³]
3 0 200
Coefficients set (for dust concentration ext) 0.00 / 1.00 / 0.00
Coefficients set (for dust concentration SI) 0.00 / 1.00 / 0.00