4.4.4 Setting the analog outputs parameters
Select the “Configuration / IO Configuration / Output Parameters” directory to set the
analog outputs.
Fig. 47: SOPAS ET menu: MCU/Configuration/IO configuration/Output Parameters
Default values see “Factory settings”, page 69
In order to output the dust concentration under standard conditions (“Conc. s.c.”
(Ext)), set the parameters for the analog outputs according to see “Setting the analog
inputs parameters”, page 74.
Field Parameter Remark
Analog Out-
puts
- General
Configuration
Output Error
current
Yes Error current is output.
No Error current is not output.
Error Current Value < Live Zero (LZ)
or > 20 mA
mA value to be output in “Malfunction” state (error case) (size
depends on connected evaluation system).
Current in
maintenance
User defined value A value to be defined is output during “Maintenance”
Value measured last The value measured last is output during “Maintenance”
Measured value
output
The current measured value is output during “Maintenance”.
Maintenance
current
Whenever possible,
value LZ
mA value to be output in “Maintenance” state