8011952/YWL2/3-0/2016-08| SICK OPERATING INSTRUCTIONS | DUSTHUNTER C200
Subject to change without notice
START-UP AND PARAMETER SETTINGS 4
4.6.3 Menu structure
Fig. 55: LC-Display menu structure
4.6.4 Parameter setting
4.6.4.1 MCU
Analog outputs / inputs
Set the MCU to “Maintenance” mode and select the “I/O Parameter” submenu.
Select the desired parameter and enter the default password “1234” using the “^”
(scrolls from 0 to 9) and/or “” (moves the cursor to the right) buttons.
Select the desired value using the “^” and/or “” buttons and write it to the device with
“Save” (confirm 2x).
Root Menu
/
Back Enter
Back Enter
Back Enter
Back Enter
Back Enter
Back Enter
1 Start Check cycle
I/O (MCU)
I/O (MCU)
I/O (MCU)
I/O (MCU)
I/O (MCU)
I/O (MCU)
I/O (MCU)
Adjustment
/io/adj
Back Enter
Back Enter
Back Enter
Back Enter
Back Enter
Back Enter
Back Enter
Back Enter
1 Operating Mode
2 Adjustment
3 I/O Diagnosis
4 I/O Parameter
1 Operating mode
2
4 I/O Parameter
Adjustment
3 I/O Diagnosis
1 AO1: 7.01mA
2 AO2: 7.02mA
1 AI1: 0.00mA
2 Ai2: 0.00mA
1 Location
2 Dresden
3 SN:08378651
4 Firmware:
5 1.0.52
1 Location
2 Dresden
3 SN:09058608
4 Firmware:
5 01.04.06
1 Operating Mode
2 Parameter
3 Diagnosis
4 Device Info
1 Operating Mode
2 Parameter
3 Diagnosis
4 Device Info
I/O (MCU)
I/O (MCU)
AO Live Values
AI Live Values
Device Info
Device Info
Sensor 1
Sensor 1
/io/
/io/
/io/diag/ao
/io/diag/ai
/io/info
//infosensor/1
/Sensor/1
/sensor/1
Back Enter
Back Enter
1 Analog output
2 Analog input
3 Device Info
1 Contam. 1.8 %
2 Span70 69,52 %
3 Zero point 0.10 %
4 LEDMon 1.895 V
5 LD_Mon 4.0 V
6 LD_I -2 mA
7 Temp 28 °C
8 Q1-4 0.03 V
I/O Diagnosis
Diagnosis
/io/diag
/sensor/1/diag
Back Enter
Back Enter
1 AO Parameter
2 AI Parameter
3 MCU Variant
1 cc2
2 cc1
3 cc0
I/O Parameter
Parameter
/io/param
//paramsensor/1
Operating Mode
Operating Mode
Operating Mode
Operating Mode
/io/mode
/sensor/1/mode
/io/mode
//modesensor/1
Menu
Opac 96.7 %
ConcSL 18.1 mg/m³
C
Ext 1.4795
R_Opac 81.8 %
Trans 3.3 %
oncEx 1.51 mg/m³
ConcEx
1.5
0.0 mg/m³ 200.0
0.0 mg/m³ 200.0
ConcSL
18.1
Menu
3
4
3
3
4
4
1): as in the operating state
2): see “MCU”, page 85
3): see “Sender/receiver unit”,
page 88
1)
2)
}
1)
}
3)