4 START-UP AND PARAMETER SETTINGS
4.6.4.2 Sender/receiver unit
The following steps are required to enter the regression coefficients:
▸ Set the sender/receiver unit to “Maintenance” and select the “Parameter” submenu.
▸ Select the desired parameter and enter the password (see “Password and operating
levels”, page 84).
▸ Set the calculated coefficients (see “Calibration for dust concentration measurement”,
page 76) using the “^” and/or “→” buttons and write to the device with “Save” (confirm
twice).