94
8011952/YWL2/3-0/2016-08| SICKOPERATING INSTRUCTIONS | DUSTHUNTER C200
Subject to change without notice
5 MAINTENANCE
5.2 Maintenance on the sender/receiver unit and reflector/scattered light receiver
Clean the outside of the sender/receiver unit and reflector/scattered light receiver in
regular intervals. Remove deposits with water or mechanically using suitable auxiliary
means (30% for warning, 40% for failure).
5.2.1 Maintenance on the sender/receiver unit
Enter the level 1 password and set the sender/receiver unit to the “Maintenance” state
see “SOPAS ET menu: DH C200/Maintenance/Maintenance”, page 59).
Loosen the knurled screws and swivel the enclosure to the side.
Check mounting flange and purge air connections for contamination and clean when
necessary.
Close the mounting flange with cover (see “Miscellaneous”, page 123).
Switch to the “Adjustment / Manual adjustment / Motor control” directory and click
“Mounting” under “Pivoted shutter sender/receiver”.
The pivoted shutter then moves to the cleaning position.
Fig. 60: SOPAS ET menu: DH C200/Adjustment/Manual Adjustment/Motor control
Remove pivoted shutter cover (1), press tension spring (7) together and take pivoted
shutter (3) off axis (6).
Carefully clean glass pane (5) (both sides), zero point reflector (2) and sender optics (4)
with an optics cloth.
NOTE:
Do not damage any device parts during maintenance work.
Do not interrupt the purge air supply.