5 MAINTENANCE
Fig. 63: SOPAS ET menu: DH C200/Diagnosis/Check values
If the measured values for contamination, zero point and span are within the allowed
ranges, save them to the device by clicking the “Refresh” button (“Check values” field); if
not, repeat cleaning and check the contamination value again by triggering a renewed
function check.
Put the pivoted shutter cover back on, remove the cover from the assembly flange,
swivel the enclosure back in and lock with the knurled screws.
Move the pivoted shutter back into the measuring position. To do this, click the
“Measurement” button in the “Adjustment / Manual adjustment / Motor control”
directory (see “SOPAS ET menu: DH C200/Adjustment/Manual Adjustment/Motor
control”, page 94).
Resume Measuring mode.
The contamination value can also be displayed on the LC display of the MCU
(initiate a function check and select the “C200/Diagnosis” menu, see “Menu
structure”, page 85).
If the contamination value does not sink below the value for warning in spite of
several cleaning processes, the device is probably defective contact SICK
Service.
If the contamination value becomes negative after cleaning (e.g. because of
scaling with contaminated optics), the measuring system must be rescaled (see
“Scaling the measuring system for transmission measurement”, page 63).