97
8011952/YWL2/3-0/2016-08| SICK OPERATING INSTRUCTIONS | DUSTHUNTER C200
Subject to change without notice
MAINTENANCE 5
5.2.2 Performing maintenance on the reflector/scattered light receiver
Set the measuring system to “Maintenance” (see “General”, page 92), loosen the
knurled screws and swivel the enclosure to the side.
Maintenance on the scattered light receiver
Push spring (2) to the side, pull bracket (3) to the front and pull scattered light receiver
optics (4) out of the tube.
Close mounting flange (5) with cover (see “Miscellaneous”, page 123).
Open tensioners (8) on the reflector/scattered light receiver for long active measuring
paths and take attachment (10) off.
Carefully clean lens of the scattered light receive optics (7) with an optics cloth.
Check purge air pipe (1), light trap (6) and O-ring (9) for deposits and clean, if required.
Fig. 64: Cleaning the optical interfaces
Scattered light receiver for short active measuring paths
Scattered light receiver for long active measuring paths


Purge air pipe
Spring
Bracket
Scattered light receiver optics
Mounting flange
Light trap
Scattered light receiver optics
Tensioner
O-ring
ß
Attachment


ß