98
8011952/YWL2/3-0/2016-08| SICKOPERATING INSTRUCTIONS | DUSTHUNTER C200
Subject to change without notice
5 MAINTENANCE
5.2.3 Maintenance on the reflector
Click the “Mounting” button in menu “DH C200/Adjustment/Manual Adjustment/Motor
control” under position Pivoted shutter reflector see “SOPAS ET menu: DH C200/
Adjustment/Manual Adjustment/Motor control”, page 98).
The pivoted shutter then moves to the cleaning position.
Fig. 65: SOPAS ET menu: DH C200/Adjustment/Manual Adjustment/Motor control
Remove pivoted shutter cover (1), loosen tension spring (6) and take pivoted shutter (2)
off axis (5).
Carefully clean glass pane (4) (both sides) and reflector optics (3) with an optics cloth.
Fig. 66: Cleaning the optical interfaces on the reflector
Lay the toothed belt on the drive axis, push pivoted shutter back onto the axis and
reinstall the tension spring.
Pivoted shutter cover
Pivoted shutter
Reflector optics
Glass pane (both sides)
Axis
Tension spring