8011952/YWL2/3-0/2016-08| SICK OPERATING INSTRUCTIONS | DUSTHUNTER C200
Subject to change without notice
MAINTENANCE 5
▸
Trigger the function check (select the “Adjustment / Function Check - Manual”
subdirectory in the “MCU“ device file and click “Start Manual Function Check”; see
“SOPAS ET menu: MCU/Adjustment/Function Check - Manual”, page 95).
▸ Check the contamination value (see “Maintenance on the sender/receiver unit and
reflector/scattered light receiver”, page 94, see “SOPAS ET menu: DH C200/Diagnosis/
Check values”, page 96).
▸ If the measured values for contamination, zero point and span are within the allowed
ranges, save them to the device by clicking the “Refresh” button (“Check values” field); if
not, repeat cleaning and check the contamination value again by triggering a renewed
function check.
▸ Put the pivoted shutter cover back on, remove the cover from the assembly flange, re-
install and fasten the scattered light receiver optics.
▸ Swivel the enclosure back again and lock with the knurled screws.
▸ Move the pivoted shutter back into the measuring position. To do this, click the
“Measurement” button under “Pivoted shutter reflector” in the “Adjustment / Manual
adjustment / Motor control” directory (see “SOPAS ET menu: DH C200/Adjustment/
Manual Adjustment/Motor control”, page 94).
▸ Resume Measuring mode.