MAINTENANCE 5
Trigger the function check (select the “Adjustment / Function Check - Manual”
subdirectory in the “MCU“ device file and click “Start Manual Function Check”; see
“SOPAS ET menu: MCU/Adjustment/Function Check - Manual”, page 95).
Check the contamination value (see “Maintenance on the sender/receiver unit and
reflector/scattered light receiver”, page 94, see “SOPAS ET menu: DH C200/Diagnosis/
Check values”, page 96).
If the measured values for contamination, zero point and span are within the allowed
ranges, save them to the device by clicking the “Refresh” button (“Check values” field); if
not, repeat cleaning and check the contamination value again by triggering a renewed
function check.
Put the pivoted shutter cover back on, remove the cover from the assembly flange, re-
install and fasten the scattered light receiver optics.
Swivel the enclosure back again and lock with the knurled screws.
Move the pivoted shutter back into the measuring position. To do this, click the
“Measurement” button under “Pivoted shutter reflector” in the “Adjustment / Manual
adjustment / Motor control” directory (see “SOPAS ET menu: DH C200/Adjustment/
Manual Adjustment/Motor control”, page 94).
Resume Measuring mode.