Process temperature specifications for Ex use
Temperature class
Ambient temperature [°C]
-40 to +40 -40 to +50 -40 to +60
T2 180 - -
T3 165 140 -
T4 100 100 80
T5 65 65 65
T6 50 50 50
EX requirements
It is required that:
● Electrical
connections are in accordance with Elex V (VO in explosion hazardous areas) and
EN60079-14 (Installing Electrical Systems in Explosion Hazardous Areas).
● The protective cover over the power supply is properly installed. For intrinsically safe circuits
the connection area can be opened.
● Appropriate cable connectors are used for the output cicuits: intrinsically safe: blue, non-
intrinsically safe: black.
● Sensor and transmitter are connected to the potential equalization. For intrinsically safe
output circuits potential equalization must be maintained along the entire connection path.
● Sensor insulation thickness is max. 100mm (only insulated sensors).
● EN60079-31 is considered for installation in areas with combustible dust.
● When protective earth (PE) is connected, no potential difference between the protective
earth (PE) and the potential equalization (PA) can exist, even during a fault condition.
Safety notes
2.3 Installation in hazardous location
SITRANS FM MAG 5000/6000 IP67
Operating Instructions, 12/2019, A5E02338368-AD 13