1. Mount hardware key on transmitter connection plate during setting of primary operating
parameters such as Q
max
., low-flow cut-off, units, approvals, etc. in connection with
commissioning or calibration. See setup menus in appendix
Transmitter menu overview
(Page 79).
2. Remove hardware key after setting up and calibrating unit.
This locks the menu structure and the selected settings.
4.5.2
Sealing of MAG 5000/6000 CT
4.5.2.1 Verification sealing
Note
The verification sealing is done at factory after initial verification.
m
n
m
nn
Under transmitter display frame
A5E34207852
'.

On bottom of transmitter
m
n
n
m
n
m
n
n
m
n
n
m
nmn
m
n
n
mnm
n
n
m
n
m
n
n
mnnmnn
On side of transmitter
Installing/Mounting
4.5 MAG 5000/6000 CT
SITRANS FM MAG 5000/6000 IP67
Operating Instructions, 12/2019, A5E02338368-AD 31