Note
For type-approved and verified MID MAG 5000/6000 CT products
The verification sealing may only be broken by an authorized person, with the acceptance and
under direction of the local authorities.
4.5.2.2 User sealing
Note
User sealing has to be done after commissioning by an authorized person.
1. Drill through marked drilling holes in terminal box and transmitter/lid.
2.
Seal the transmitter on both sides with either one or two wires, as shown below.
Compact version sealed with
one wire
Remote version sealed with
one wire
Compact version sealed with
two wires
4.5.3 Installation conditions
4.5.3.1
MI-001
MAG 5000/6000 CT together with MAG 5100 W (7ME652) are approved for MI-001 under the
following installation conditions.
DN 50 to 300 (2" to 12") in any orientation
DN 350 to 1200 (14" to 48") only in horizontal installation
Compact or remote with max. 500 m (1640 ft.) cable
Power supply 115 to 230 V AC and 12 to 24 V DC
Installing/Mounting
4.5 MAG 5000/6000 CT
SITRANS FM MAG 5000/6000 IP67
32 Operating Instructions, 12/2019, A5E02338368-AD