Power on transmitter
1 Flashing error triangles? Yes ⇒ Check error table
No ⇒ 1.2
1.2 Output and display readings OK? Yes ⇒ 1.2.1
No ⇒ 1.2.2
1.2.1 Transmitter OK Check application
Check installation/sensor/earthing connec‐
tion etc.
1.2.2 Check cables/connections
Check connection board
Check pins in transmitter multiplug
OK ⇒ 1.2.1
Not OK ⇒ correct fault
2 Check cables/connections
Check connection board
Check pins in transmitter multiplug
OK ⇒ 2.1
Not OK ⇒ Correct fault
2.1 Output readings OK? Yes ⇒ 2.1.1
No ⇒ 2.1.2.
2.1.1 Display defective Replace display
2.1.2 Transmitter defective Replace transmitter
Note
Sensor check list
Check list for sensors are included in the relevant sensor Operating Instructions.
8.2 Technical support
If
you have any technical questions about the device described in these Operating Instructions
and do not find the right answers, you can contact Customer Support:
Via the Internet using the Support Request:
Support request (
http://www.siemens.com/automation/support-request)
Via Phone:
Europe: +49 (0)911 895 7222
America: +1 423 262 5710
Asia-Pacific: +86 10 6475 7575
Further information about our technical support is available on the Internet at
Technical support (http://support.automation.siemens.com/WW/view/en/16604318)
Service & Support on the Internet
In addition to our documentation, we offer a comprehensive knowledge base online on the
Internet at:
Service and support (http://www.siemens.com/automation/service&support)
Service and maintenance
8.2 Technical support
SITRANS FM MAG 5000/6000 IP67
58 Operating Instructions, 12/2019, A5E02338368-AD

Table of Contents

Related product manuals