Note
Uncertainties for MAG 5000/6000 CT systems under MID requirements
± 5% for flowrate between Q1 and Q2
± 2% for flowrate between Q2 and Q4 if media temperature
≤ 30°C
More detailed installation conditions can be found in the certificate.
Reference conditions
(ISO 9104 and DIN/EN 29104)
A calibration certificate is shipped with every sensor and calibration data is stored in
SENSORPROM memory unit.
Table 10-1
Reference conditions for sensor calibration
Medium temperature 20°C ± 5°C (68°F ± 9°C)
Ambient temperature 20°C ± 5°C (68°F ± 9°C)
Supply voltage U
n
± 1%
Warming-up time 30 minutes
Technical data
10.2 Accuracy
SITRANS FM MAG 5000/6000 IP67
Operating Instructions, 12/2019, A5E02338368-AD 71