Reminder 2 before Lifetime (Demanded) (4.3.6.)
If Remaining Lifetime (4.3.3.) is equal to or less than this value, the device generates a
Maintenance Demanded reminder.
Values
Range: 0 to Lifetime (Expected) (4.3.1.)
1. Modify values as required.
2. Set Activation of Reminders (4.3.4.) to the desired option.
Maintenance Status (4.3.7.)
Indicates which level of maintenance reminder is active.
In SIMATIC PDM, open the menu View – Device Status, click on the Maintenance tab, and
check the
Sensor Lifetime Status window.
Acknowledge Status (4.3.8.)
Indicates which level of maintenance reminder has been acknowledged.
In SIMATIC PDM, open the menu View – Device Status, click on the Maintenance tab, and
check the
Sensor Lifetime Status window.
Acknowledge (4.3.9.)
Acknowledges the current maintenance reminder.
To acknowledge a reminder via SIMATIC PDM:
1. Open the menu View – Device Status and click on the Maintenance tab.
2. In the Sensor Lifetime section, click on Acknowledge Warnings.
To acknowledge a reminder via the handheld programmer:
1. Press RIGHT arrow twice to open parameter view and activate Edit Mode.
2. Press RIGHT arrow to acknowledge the reminder.